Olympus VISERA クイック・リファレンス・マニュアル
VISERA UHD 4K Camera System Quick Reference Guide
Disclaimer: Information provided in this guide is not exhaustive. Please refer to the complete Instructions For Use for each
item of equipment which should be thoroughly read and understood prior to handling, use and storage of the equipment. 
Camera Head
Attach the camera head to the telescope by rotating
the endoscope mount as below.
Endoscope
Rotate
lock ring
Eyepiece
Endoscope mount
Releasing the camera head mount secures the telescope
in place.
Issue 1
05/02/2021
White Balancing
Once the camera head, telescope and light guide cable
are connected, a white balance must be performed.
Perform a white balance by surrounding the tip of the
telescope in a white swab, ensuring it is fully enclosed.
Press and hold the white balance button on the
processor until 'White Balance Complete' appears
on the monitor.
White balance window
The white balance status changes to completed.
White balance window
High Intensity Mode
Brighter examination light is available when connecting an
endoscope that has an insertion tube outer diameter of
4.1 mm or greater and a compatible Olympus light guide
cable by pressing the intensity mode button.
High intensity mode
indicator
Intensity mode button
Warning:
Do not use the high intensity mode when
connecting to an endoscope that has an insertion tube
outer diameter of 4.1 mm or less. The endoscope
becomes hot while the lamp is lit up and could
cause operator burn.
Caution:
Do not touch the endoscope's distal end and
light guide connector while the examination lamp is lit up.
The endoscope becomes hot while the examination lamp
is lit up and could cause operator burn.