顕微鏡 Olympus BH-2 BHSUのPDF 完全な分解、洗浄、再組み立てをオンラインで閲覧またはダウンロードできます。Olympus BH-2 BHSU 23 ページ。 Sliding focus block

Olympus BH-2 BHSU 完全な分解、洗浄、再組み立て
Figure 24 – Clean grease from side groove in focus block 
Remove the Guide Rails from the Guide Block 
Remove the two short guide rails from the side groove 
in the guide block (see Figure 25). 
Figure 25 – Remove guide rails from the guide block 
Use  a  suitable  solvent  (e.g.,  acetone)  to  remove  any 
grease  from  two  short  guide  rails  (see 
from the side groove in the guide block (see Figure 26).  
Figure 26 – Clean grease from side groove in guide block 
Clean the Cylindrical Rollers 
Soak  the  cylindrical  rollers  in  a  suitable  solvent  (e.g., 
acetone)  to  remove  all  grease  from  the  rollers  (see 
Figure 27). 
Complete Teardown, Cleaning, and Reassembly of the Olympus BH‐2 Sliding Focus Block  
 
 
Reassemble the Sliding Focus Block 
Reassemble  the  sliding  focus  block  mechanism  per  the 
following procedure. 
Reinstall the Guide Rails into the Arm Casting 
Per  the  procedure  described  in 
document,  reinstall  the  four  long  guide  rails  into  the 
two side grooves of the arm casting (see Figure 28) and 
remove  any  exposed  grease  from  the  four  guide  rails 
(see Figure 29 and Figure 30).   
 
Figure 
19)  and 
 
Figure 27 – Soak the rollers in solvent to remove grease 
Figure 28 – Reinstall the long guide rails into the arm 
Figure 29 – Scrub grease from guide rails and arm 
Appendix 
1  of  this 
Revision 2 