- ページ 9

顕微鏡 Olympus BH2-5REのPDF 完全な分解、洗浄、再組み立てをオンラインで閲覧またはダウンロードできます。Olympus BH2-5RE 18 ページ。 Modular revolving nosepiece

Olympus BH2-5RE 完全な分解、洗浄、再組み立て
Figure 13 – Remove the pivot-adjustment screw

Remove the Center-Pivot Bearing Ball

Use a tweezers to remove the ¼" bearing ball from the
center bore in the stationary base (see
Figure 14 – Remove the center-pivot bearing ball

Remove the Mechanical Detent Stop

The next step is to remove the mechanical detent stop
from the stationary base of the turret assembly. Use a
suitable JIS screwdriver to remove the two M2X4 pan-
head screws securing this stop to the stationary base
(see
Figure
15). These screws are staked in place with
adhesive and can be surprisingly stubborn to remove, so
be sure to use the proper JIS screwdriver to prevent
damaging the screw heads. It might also be helpful to
heat the screws with a heat gun before loosening them,
but do not melt the black knurled-rubber grip ring in the
process.
Complete Teardown, Cleaning, and Reassembly of the Olympus BH2-5RE Modular Revolving Nosepiece
Figure
14).
Figure 15 – Remove the screws securing the detent stop
Remove the loose mechanical detent stop from the
stationary base (see
Figure
Figure 16 – Remove the loose mechanical detent stop

Remove the Knurled-Rubber Grip Ring

Carefully remove the black knurled-rubber grip ring
from the outer perimeter of the revolving turret (see
Figure
17), being careful not to stretch or damage it in
the process. This ring is held onto the turret with a few
spots of adhesive, so proceed carefully here.
Figure 17 – Remove the knurled-rubber grip ring
16).
Page 9 of 18
Revision 3