- ページ 10

顕微鏡 Olympus BH2-CH SubstageのPDF 組み立てをオンラインで閲覧またはダウンロードできます。Olympus BH2-CH Substage 26 ページ。

Olympus BH2-CH Substage 組み立て
Figure 16 – Remove screws and washers securing rack
Remove the loose condenser-height rack from the face
of the dovetail slide (see
Figure 17 – Remove the condenser-height rack
Look for a brass shim either on the back of the
condenser-height rack or on the face of the dovetail
slide (see
Figure
18).
Figure 18 – Brass shim for the condenser-height rack
Remove the brass shim from the back of the condenser-
height rack or from the face of the dovetail slide,
whichever is applicable (see
Complete Teardown, Cleaning/Lubrication, and Reassembly of the Olympus BH2-CH Substage Assembly
Figure
17).
Figure
19).
Figure 19 – Remove brass shim for condenser-height rack

Disassemble the Condenser-Height Stop

Hold the M5x12 condenser-height stop pin down with
your finger, and use a 1.5mm Allen wrench or driver to
loosen the M3x4 set screw holding the stop pin in place.
While holding your finger on top of the stop pin, allow
the stop pin to rise in order to alleviate the compression
on the internal stop spring (see
Figure 20 – Loosen set screw while holding the stop pin
Remove the M5x12 stop pin from the bore in the
condenser slide (see
Figure
Figure 21 – Remove the loose stop pin from the bore
Remove the stop-pin spring from the bore in the
condenser slide (see
Figure
Figure 20).
21).
22).
Revision 1
Page 10 of 26