- ページ 12
顕微鏡 Olympus BHSUのPDF 完全な分解、洗浄、再組み立てをオンラインで閲覧またはダウンロードできます。Olympus BHSU 19 ページ。 Reverse nosepiece
Figure 27 – Reinstall stationary base into revolving turret
Apply a Ring of Grease for the Perimeter Balls
The next step is to apply grease into the ring in which
the perimeter bearing balls will be placed. Carefully
apply a ring of grease (item 8 of
channel formed between the outer perimeter of the
stationary base and the inner perimeter of the revolving
turret (see
Figure
28). Do not apply too much grease
here, to minimize the squeeze-out that will occur when
the threaded retaining ring is reinstalled to hold the
stationary base into the recess of the revolving turret.
Figure 28 – Apply ring of fresh grease for the bearing balls
Reinstall the Perimeter Bearing Balls
Use tweezers to carefully set the 3/32" bearing balls
(there are 80 of them) into the grease ring (see
29), placing the bearing balls as close together as
possible as you proceed.
Complete Teardown, Cleaning, and Reassembly of the Olympus BHSU/BHTU Reverse Nosepiece Assembly
Appendix
1) into the
Figure
Figure 29 – Place the bearing balls into the ring of grease
Reinstall the Threaded Retaining Ring
Carefully engage the threads of the threaded retaining
ring with the threads in the revolving turret (see
30).
Figure 30 – Engage the threaded retaining ring
Use a suitable lens-spanner tool (item 6 of
in a pair of opposing tooling holes to tighten the
2
threaded retaining ring
Figure 31 – Tighten the threaded retaining ring
2
Or use a center punch or nailset tool to snug the threaded retaining ring by
placing the tip of the tool into one of the four tooling holes and lightly
tapping the tool with a small hammer or mallet to drive the threaded
retaining ring clockwise.
Appendix
(see
Figure
31).
Revision 3
Page 12 of 19
Figure
1)