- ページ 33

顕微鏡 Olympus BX-RFAAのPDF 取扱説明書をオンラインで閲覧またはダウンロードできます。Olympus BX-RFAA 42 ページ。

9-2 Detailed Assembly Procedures

1
Attaching the Revolving Nosepiece (Motorized Revolving Nosepiece)
}With certain series of motorized revolving nosepiece products, the connector of the BX-RFAA illuminator should be
unlocked before proceeding to the attaching operation as described below.
30
@
Fig. 10
U-D6REM/5BDREM
These revolving nosepieces can be installed in the same way as manual
revolving nosepieces. For installation, push the revolving nosepiece as far
as it will go to connect it to the motorized revolving nosepiece connector.
U-D6REO/P5REMC
Remove the screw @ which has been clamped at the factory using a
Phillips screwdriver. This unlocks the connector so the revolving nose-
piece can be installed in the same way as the above.
To re-lock the connector:
After installing the motorized revolving nosepiece, clamp the screw @.
Note that, in this case, other motorized revolving nosepieces than the
combined motorized revolving nosepiece may sometimes be unable to
be connected due to the lack of compatibility with other motorized re-
volving nosepieces.