- ページ 22
顕微鏡 Olympus BX3M-CBのPDF 取扱説明書をオンラインで閲覧またはダウンロードできます。Olympus BX3M-CB 44 ページ。 Control box/fm control box
: Turns ON.
: Blinks.
Number
of beeps
3 times
None
None
18
:Turns OFF.
: Varies depending on state.
Phenomena
CB
HS
–
–
–
Cause
MIX
The motorized
Set the main switch to
nosepiece is not
connect the motorized nosepiece
connected.
again.
–
The motorized
Contact Olympus.
nosepiece or its
cable is damaged.
The hand switch
Set the main switch to
for motorized
and connect the hand switch for
nosepiece (BX3M-
motorized nosepiece (BX3M-HSRE).
HSRE) is not
If the indicator4 on the side or the
back of the control box does not
connected.
turn ON even though the hand
–
switch for motorized nosepiece is
connected, contact Olympus.
The hand switch
Contact Olympus.
for motorized
nosepiece (BX3M-
HSRE) is damaged.
The hand switch
Contact Olympus.
for motorized
nosepiece
–
(BX3M-HSRE) or
the motorized
nosepiece is
damaged.
Remedy
Page
(OFF) and
9, 27
–
(OFF)
9, 27
–
–