- ページ 6

顕微鏡 Olympus BX46のPDF 取扱説明書をオンラインで閲覧またはダウンロードできます。Olympus BX46 44 ページ。 Clinical microscope

Olympus BX46 取扱説明書
Restrictions in Use
When all combinations (@ to ³) below are satisfied at the same time, the microscope may fall if tilted larger than a certain
degree, comparing to other combinations. Please mount the camera port of the intermediate observation tube toward
diagonally backward 45° to prevent from falling.
@ The observation tube is either the trinocular tube, U-TTBI, U-ETBI or U-TTLBI.
² The intermediate observation tube is either U-TRUS or U-DP.
³ The camera heavier than 0.4kg is mounted on the intermediate observation tube described in ².

Safety Symbols

The following symbols are found on the microscope. Study the meaning of the symbols and always use the equipment
in the safest possible manner.
Symbol
Caution engraving/label
Caution engraving and label are placed at parts where special precaution is required when handling and using the
microscope. Always heed the cautions.
If a caution engraving or label is dirty or peeled off, contact Olympus for the replacement or other inquiry.
1
Getting Ready
Fig. 2
2
Indicates that the surface becomes hot, and should not be touched with bare hands.
Indicates a non-specific general hazard. Follow the description given after this symbol or in
instruction manual.
Indicates that the main switch is ON.
Indicates that the main switch is OFF.
Caution engraving
Halogen lamp socket
position
[Caution against high temperature]
Caution label
Rear panel
position
[Caution against high temperature]
1. A microscope is a precision instrument. Handle it with care and avoid
subjecting it to sudden or severe impact. Also note that the microscope
does not have a waterproof construction.
2. Do not use the microscope where it is subjected to direct sunlight, high
temperature and humidity, dust or vibrations. (For the operating conditions,
refer to Chapter 8, "SPECIFICATIONS".)
3. When moving the microscope, remove the observation and specimen
and carefully carry the microscope by holding the grasping part on the
rear of the arm and the front part of the base as shown in Fig. 2.
CAUTION
· Even when moving the microscope along the desktop surface, be sure
· When the microscope needs to be packaged for forwarding to a remote
4. When the accessories of this product are used, make sure that the system
height does not exceed 1 meter. Otherwise, the stability of the device will
be lost.
5. When an arm, illuminator, intermediate attachment, observation tube and
camera are to be attached to the upper part of the microscope frame,
make sure that their total weight does not exceed 7 kg.
Explanation
To prevent damage to the microscope, never hold it by the stage
or coarse/fine adjustment knob.
to remove the specimen in advance.
location, contact Olympus.
(Fig. 2)