- ページ 8

顕微鏡 Olympus BX51-PのPDF 取扱説明書をオンラインで閲覧またはダウンロードできます。Olympus BX51-P 36 ページ。 Polarizing microscope
Olympus BX51-P にも: インフォメーション (6 ページ), 取扱説明書 (36 ページ)

Olympus BX51-P 取扱説明書
³
Fig. 3
5
2
Mounting the Condenser (U-POC-2)
1. Turn the coarse adjustment knob @ to raise the stage to its upper limit.
²
2. Turn the condenser height adjustment knob ² to lower the condenser
holder to its lower limit.
3. Loosen the condenser clamping screw ³.
4. Position the condenser with the scale markings in front, insert it into the
substage fork as far as it will go.
Align the positioning pin at the back of the condenser with the groove in
@
the substage fork.
CAUTION
Swing the top lens out of the way before inserting the condenser.
5. Tighten the condenser clamping screw, then raise the condenser to its
upper limit.
(Fig. 3)