- ページ 33

顕微鏡 Olympus BXC-CBBのPDF 取扱説明書をオンラインで閲覧またはダウンロードできます。Olympus BXC-CBB 41 ページ。 Modular microscope assemblies hardware

4.
At this time, if the center of the aperture diaphragm is shifted, adjust it by rotating the aperture diaphragm centering
screws (3 positions) with the Allen screwdriver.

4.4 Using the interlocked ND filter for BXC-RLI

TIP:
The ND filter is interlocked with the brightfield (BF) light path switching and makes it possible to reduce glare when the
darkfield (DF) path is switched to the brightfield (BF) path.
Removing the Interlocked ND Filter
TIP:
The ND filter has been interlocked at the factory. If brightness is not enough during brightfield, DIC or other observation,
remove the ND filter as described below.
1.
Switch the observation path to DF, then loosen the cover clamping screw (A) of the interlocked ND filter using the
Allen screwdriver provided with BXC-RLI and remove the cover.
29
4. Operation procedures
4.4 Using the interlocked ND filter for BXC-RLI