- ページ 13

顕微鏡 Olympus CH10のPDF 修理マニュアルをオンラインで閲覧またはダウンロードできます。Olympus CH10 31 ページ。 Biological microscope

Olympus CH10 修理マニュアル
CH10/CH20
±
GEAR ASS'Y
1 0
±
COARSE ADJ.
1 3
KNOB-L
±
FINE ADJ.
1 4
KNOB
±
FINE ADJ.
1 5
SHAFT
±
RING
1 6
±
FINE ADJ.
1 7
GEAR
±
RING
1 8
C. DISASSEMBLY AND ASSEMBLY PROCEDURES
OT2008
Insert the gear ass'y Š into the groove of
coarse adj.shaft
A at the right side with the gear inside of
the cam
OT1028
Apply adhesive on the screw thread of fine
adj. shaft
Apply grease on the sliding surface.
Press in to 101 mm from the end of the
fine adj. shaft
page.
C-3
±
and engage the big gear
1 1
±
.
1 2
±
.
1 5
±
. See Figure-A on previous
1 5