- ページ 29

顕微鏡 Olympus CK40M FRAMEのPDF 修理マニュアルをオンラインで閲覧またはダウンロードできます。Olympus CK40M FRAME 38 ページ。 Compact inverted metallurgical microscope

Olympus CK40M FRAME 修理マニュアル
CK40M
5.

CK4M-MA illumination center adjustment

5-1
Set the following jigs as illustrated below to adjust the CK4M-MA illumination center.
• Eyepiece-----KN0029: Centering telescope
• Other --------BKN0003: Standard tube, stage insert plate, mirror
The mirror side
should face down.
AB 5X15SA
ANU 3X4SA (3 pcs.)
C. REPAIR PROCEDURE
AB 5X15SA
Mirror
Stage insert plate
C-8
Centering telescope
(KN0029)
Standard tube
(BKN0003)
Insert the
slide glass.
Slider mount
(ANU 3X4SA; 2 pcs.)