- ページ 35

顕微鏡 Olympus CKX41のPDF 修理マニュアルをオンラインで閲覧またはダウンロードできます。Olympus CKX41 43 ページ。 Reflected fluorescence system
Olympus CKX41 にも: 概要 (7 ページ), パンフレット&スペック (4 ページ), 取扱説明書 (36 ページ)

Olympus CKX41 修理マニュアル
CKX31/CKX41

3. Adjustment procedure (CKX41)

Preparation for optical adjustment
Adjusting parfocality
Adjusting revolving axis
Adjusting left / right optical axis
Adjusting absolute optical axis
(center of standard objective)
Checking exit pupil center
Checking stage tilt alignment
Adjusting moving force (tension)
*In adjustment of revoving axis and left/right optical axis , it is nesessary to use CK40-MVR (and
supplied slide-glass holder ) when moving a specimen. If not available, move it by hand.
D. REPAIR PROCEDURE
See D-9.
See D-10.(adjustment in binocular tube)
See D-11 to D-12.
Jigs : See D-10.
Procedure: See E-3 to E-6
Jigs : See D-10.
Procedure: See E-8 to E-9
Jigs : See D-9 and D-10.
Procedure: See E-10 to E-11
Jigs : See D-9 and D-10.
Procedure: See D-14 (6-2) in CK30/40
See D-17 in CK30/40 repair manual.
See C-6 to C-7.
D-8
in CH3-BI45 repair manual.
in CH3-BI45 repair manual.
in CH3-BI45 repair manual.
repair manual.