MICROSCOPE CONTROL FROM COMPUTER
}Connecting the MX61/61L microscope with a host computer using an RS-232C cable makes serial communications
available for computer control of the microscope.
@
3
Connection of Connectors and Pin Assign
27
Fig. 37
1
RS-232C Communication Parameters
}The RS-232C communication parameters are fixed as shown below.
Transfer rate
Data bits
Parity
Stop bits
Flow control
Terminator
(Line feed code)
2
Connection of the RS-232C Cable
This connector is only for connection of a computer that corresponds
to the requirements of IEC60950.
# The RS-232C cable used must be a straight cable that is commercially
available. (The cable should be equipped with two D-sub 9P female-
female connectors on the extremities.)
· Be sure to set the main switches of the microscope system and host
computer to "
" (OFF) before connecting the RS-232C cable @.
19,200 [bps]
8 [bits]
Even
2 [bits]
None
(CTS control applied only after power ON)
CR+LF (ASCII codes 0x0D, 0x0A)
(Fig. 37)