Rotate the zooming knob c to the lowest magnification, , look
only into the left eyepiece, and focus the specimen by rotating the
left diopter adjustment ring d instead of the coarse and fine focus
adjustment knobs.
Note (or memorize) the diopter readings of the left and right
TIP
eyepiece scales so that they can be duplicated quickly in the
next observation.
3
Using the Eye Shades
When Wearing Eyeglasses
Use with eye shades in their normal folded-down position. (This will
prevent eyeglasses from being scratched by the eyepiece.)
When Not Wearing Eyeglasses
Extend the folded eyeshades in the direction of the arrow. This makes
observation easier by preventing the inverse incidence of light from
between the eyepiece and your eyes.
4
Mounting the Eyepiece Micrometer Disk
· A variety of eyepiece micrometer disks can be inserted into
TIP
the WHN10XH, WHSZ15X-H and WHSZ20X-H eyepieces.
Use 24 mm dia. x 1.5 mm thick micrometer disks.
Following figure, turn the built-in micrometer-mounting frame
b to the arrow direction to remove it from the eyepiece and
place a micrometer disk a into the mounting frame so that
the surface with the model indication faces downward.
· The micrometer-mounting frame may be tight for certain
micrometer disks.
In this case, turn the frame by holding the circumference
with a light, uniform force or by applying the frame against a
rubber sheet. Do not grasp the frame with a strong force, as
this may deform the frame and make it harder to remove it.
Re-attach the micrometer mounting frame in the original
position.
· Be careful not to touch the lens surface with your finger.
The WHSZ20X-H eyepiece is designed to apply a
Note
magnification to the focusing plane of the micrometer disk.
As the magnification coefficient is 1.35, be sure to perform
magnification compensation when using this eyepiece in
measurement.
When the micrometer disk is engaged in their light path,
the light path length will extend and the diopter scale may
deviate toward the + direction from the normal indication.
However, this does not pose any problem in actual
observation.
When the eyepiece micrometer disk is not used, wrap it in a
TIP
clean sheet of soft paper before storage.
SZX16
11