4-4 Differences in Operating Procedures of Software BX2-BSW
}The BX2-BSW software has been designed for controlling the motorized operations of the BX61 microscope system.
When the illuminator is controlled from the BX2-BSW, some operations may differ from the information shown in the
operation screens, tutorial and help of the software as described below.
1-2 Setting Procedure
1-2-2 Initial setting
2. Motorized unit setting (Device Type)
· Frame
· Reflected light illumination
· Filter wheel
· Autofocusing unit
· Focus adjustment knob unit
(Note)
Text "enabled when the focus adjustment knob unit is installed" should be deleted.
3. Observation method setting (Microscopy Tab)
· Check only "EPI BF" (reflected light brightfield observation) and "EPI DF" (reflected light darkfield observation).
4. Mirror unit setting (Mirror Unit Tab)
· Enter "BF" or "DF".
6. Turret optical device and top lens setting (Condenser Tab)
· Do not register anything.
7. Filter setting (Filter Tab)
· The U-FWT cannot be used.
1-2-4 Button function allocation
1. Button selection
Main Unit Tab
· Allocation to all of buttons 1 to 8 is possible with the PC screen operation.
· Even when button functions are allocated on the main unit, standalone operation on the main unit is impossible
(because the buttons are not present).
Other caution
· The microscope illustrations in the operation screens show the BX61.
BX2-BSW Help
Focus adjustment range setting
· The relationship between the objective moving direction and the Near/Far Limits is identical to that of the BX61WI.
· The height values entered as the Near and Far Limits cannot exceed the lower hardware limit (which is adjustable).
16
Enter "BX61F" or "BX61WI". (Any can be entered without problem.)
BX-RLAA
U-FWR (The U-FWT cannot be used.)
U-AFA1M
U-FH