Fig. 11
Connecting the Power Supply for Mercury/
6
Xenon Burner
}For the basic connections, refer to the instruction manual provided with
the power supply for mercury/xenon burner.
The following description pertains to the settings required for use of this
unit and the method of attaching the conversion cable.
1. Set the 75 W/150 W switch to 150 W.
2. Use the same output setting as the factory-shipped setting.
3. Use the conversion cable @ for the connection of the lamp housing.
Connect the interlock cables ² to the positions shown in the figure.
· Red cable: To INTERLOCK (+).
· Blue cable: To INTERLOCK (-).
7
Connecting Other Cables
1. When connecting a module equipped with a main switch, set it to "
(OFF) before connection or disconnection.
2. For safety, the connection of the power cord into the power outlet should
always be the last connected cable.
3. Be sure to connect each cable firmly.
U-UVF248
(Fig. 11)
"
16