8-2 Detailed Assembly Procedures
1
Unlocking Transportation Lock
2
Installing the scanning stage with ultrasonic (When using BX63)
1. Place BX3-SSU over BX63 so that the respective screw numbers (see the Fig. 6) of BX3-SSU and BX63 are overlapped.
2. Tighten the screw hole @ with the clamping screw (attached with BX3-SSU) from the upper side of BX3-SSU. Use Allen
wrench (40x70mm) attached with BX63 for tightening.
3. Tighten the screw holes 2 and 3 with the clamping screw (attached with BX63) from the lower side of the stage fixing
pillar of BX63. Use Allen wrench (40x70mm) attached with BX63 for tightening.
Front side
Clamping screw (attached with BX3-SSU)
}Tighten with Allen wrench (40x70mm).
13
Downloaded from
ManualsNet.com
1
Fig. 11
Clamping screw (attached with BX3-SSU)
}Tighten with Allen wrench (40x70mm).
Upper side of BX3-SSU
3
search engine
Remove the screw of fixtures @ (4 positions) with the commercially
available Allen wrench.
} As fixtures and screws will be used in transportation, be sure to keep
them in a safe place.
2
1
Front side
2
BX63
Fig. 12
Allen wrench (AxBmm)
A
3
Lower side of BX3-SSU
1
(Fig. 11)
(Fig. 12)
B