1. Inspection of the unit and the accessories
No.
010
Accompanying documents are present and legible:
- Instructions for Use
- EMC-Guidelines
- final test report (incl. test protocols according to the final test report)
- optional (depending on the country): medical device information ("Medizinproduktebuch")
020
Labels present and legible:
SN: 3580-1001
Samples of type plates: Details of the manufacturer, supply voltage, supply frequency, moisture
protection class (IP), reference number, serial number (e.g. 1234W34-567), power input, power
output, and output frequency.
Visible on the back of the unit (see fig. 1)
© Celon AG
,
Issue date: 01.09.2008
SCI.991007-1.4
Safety Check
Instruction.doc
Safety Check Instruction
Sample of symbol:
Sample of symbol:
3580A34-567
Procedure
"Unit is defibrillator-safe"
Visible on the front panel (see fig. 2)
"Caution! Read the Instructions for Use!"
Visible on the front panel (see fig. 2)
1234W34-567
1234W34-567
Page 2 / 9