Dry the Air/Water Channel
1.
Turn the Power Switch ON "
2.
Close the irrigation port of the endoscope with your finger. Press the AW
Channel Cleaning Adapter button and release it. Repeat this procedure at
intervals of approximately 30 seconds until no more water or 70% ethyl or
isopropyl alcohol exit the Distal End of the endoscope (see Figure 4.2).
3.
Turn the Power Switch OFF "
4.
Remove the Light Guide Connector from the Output Socket.
Figure 4.2
MAINTENANCE UNIT MU-1
".
".
Chapter 4 Operation