4 Remedial actions in case of emergency
As the measures to be taken until the spare equipment is available, this instrument provides the following remedial
actions.
In case the arm cannot be locked
1 Flip open the switch cover of this instrument and
press the arm locking switch to lock the arm.
2 When the view field movement stop switch is pressed, movement of the microscope section by pressing the
scope switch C may become heavy.
In case movement by joystick control cannot be stopped
1 Flip open the switch cover of this instrument and
press the view field movement stop switch to stop the
joystick operation.
2 Press the scope switch C and move the microscope section.
In case focus does not work
Press the scope switch C and move the microscope section to focus on the observed region.
In case zoom does not work
1 Pull out the manual zoom knob and confirm that the status indication
line (orange) is visible. A message is displayed on the observation
monitor indicating that the manual zoom knob has been pulled out.
2 Turn the manual zoom knob to control zoom.
– 11 –
Switch cover
Arm locking switch
Switch cover
View field movement
stop switch
Status indication
line (orange)
Manual
zoom knob