Keep adequately away from cords of electric scalpel (scalpel holder, scalpel cord and
patient plate cord), scalpel main body and cords, main body of other electromedical
equipment.
Do not loop cords of electric scalpel (scalpel holder, scalpel cord and patient plate
cord). Do not contact these cords with the patient or metallic portion of an operating
table, etc.
The PSD-10 should not be used in conjunction with electrical apparatus whose safety
against leakage current is not guaranteed.
Radio-frequency and spark discharge "noise" may adversely affect other electro-medical
equipment being used in the proximity.
Do not apply the electric scalpel to the patient who has a pacemaker in his body.
The use of a radio-frequency noise filter in monitor electrode lines may reduce interference.
Patient monitor electrodes must be placed as far away as possible from the operative
field.
Needle type electrodes should not be used as they increase the risk of burns.
To prevent instrument damage, the active electrode and the patient plate should not
be short-circuited.
Do not activate the electrode for an extended time when it is not touching the target
tissue.
The PSD-10 should be used in accordance with the operating environment specified
in 1 – 2 Main Specifications; otherwise improper performance, compromised safety or
instrument damage may result.
Repairs should be carried out only by an Olympus Service Center.