Olympus CKX41 修理マニュアル
CKX31/CKX41

2. Adjustment procedure (CKX31)

Preparation for optical adjustment
Adjusting parfocality
Adjusting revolving axis
Adjusting left / right optical axis
Adjusting absolute optical axis
(center of standard objective)
Checking exit pupil center
Checking stage tilt alignment
Adjusting moving force (tension)
*In adjustment of revoving axis and left/right optical axis , it is nesessary to use CK40-MVR (and
supplied slide-glass holder ) when moving a specimen. If not available, move it by hand.
D. REPAIR PROCEDURE
See D-4.
See D-5 to D-6.
Jigs : See D-4
Procedure: See D-3 (3-2, 3-3) to D-7
Jigs : See D-4
Procedure: See D-8 (4-2, 4-3) to D-10
Jigs : See D-4
Procedure: See D-11 (5-2) to D-13
Jigs : See D-4
Procedure: See D-14 (6-2) in CK30/40
See D-17 in CK30/40 repair manual.
See C-5 to C-6.
D-3
in CK30/40 repair manual
in CK30/40 repair manual
in CK30/40 repair manual
repair manual