- ページ 26
ラボラトリー機器 Olympus MX61AのPDF マニュアルをオンラインで閲覧またはダウンロードできます。Olympus MX61A 48 ページ。 Motorized 300 mm compatible semiconductor/fpd inspection microscope
@
²
Fig. 25
23
14
Attaching the AF Control Board
# Be sure to set the main switch of the BX-UCB control box to "
before mounting the U-ZPCB or U-AFA2M-CB auto focusing control
board in it.
CAUTION
If any Z-focus control board is mounted in the BX-UCB,
remove it before proceeding. Otherwise, a failure may
result.
1. Loosen the six clamping knobs from the two option slot covers on the
rear of the BX-UCB @, and remove the option slot covers together with
the clamping knobs.
}Attach the removed clamping knobs on the AF control board ² so that it
can be inserted or removed by holding the clamping knobs.
2. Aligning the orientations of the connector of the AF control board ² with
that inside the BX-UCB, insert the AF control board into the BX-UCB
along the board rails.
Push the AF control board all the way until the connectors are connected
firmly.
3. Clamp the AF control board ² with the clamping knobs above.
Also attach one of the option slot covers in the same way.
}Retain the unused option slot cover in a safe place.
}When using the U-DICRM motorized DIC slider, it is also necessary to
mount the U-DICPCB DIC/DUV option board now (for details, see the
instruction manual for the U-DICRM).
(Fig. 25)
" (OFF)