CONTROLS
Excitation filter slider (Page 8/16)
Filter turret (Page 5)
Brightness iris diaphragm
lever (Page 5)
Shield plate
Must be used to protect
your eyes during U-excited
observation.
Unless the following precautions are observed, the UV rays will hit the specimen
surface and cause a hazardous situation.
*1. Combinations of filter units for the specified excitation and barrier filters must always
be attached along a single line as shown in the illustration above. Also incorrect
combination will cause a hazardous situation.
*2. Always attach a dummy unit whenever no filter unit is used.
Burner ON LED
Hour counter
Main switch
*
1
*
2
Power Supply Unit
(for 100W mercury burner)
Collector lens focus adjustment knob (Page 14)
One of the optional 32ND6,
32ND12, 32ND25 and 32ND50
filters can be mounted.
Pulled out: ND IN.
Pushed in: ND OUT.
Barrier filter slider (Page 8/16)
Up to 5 sliders can be mounted.
· SZX-FU
· SZX-FBV
· SZX-FCFPA
· SZX-FGFP
· SZX-FGFPA
· SZX-FG
· SZX-FREP1
· SZX-FREP2
AC receptacle
SZX–RFL3
Mirror focus adjustment
screw (rear panel) (Page 15)
Lam centering knobs
(Page 14)
Shutter slider (Page 5)
Pulled out: Shutter IN.
Pushed in: Shutter OUT.
ND filter slider (Page 5/15)
*
1
*
2
Output connector
4