SYSTEM STARTUP PROCEDURE

}The system can be started using dedicated software (analySIS FIVE or MX2-BSW).
For details, refer to the instruction manual for the software.
10
³
Fig. 5
Installing the AF Control Board
# Be sure to set the main switch on the BX-UCB control box @ to "
(OFF) before proceeding to installing the U-AFA2M-CB AF control
board.
1. Loosen the six clamping knobs on the two option slot covers on the rear
of the BX-UCB and remove the covers and knobs.
}To stabilize the installation operation, attach two of the removed clamping
knobs on the AF control board ².
2. Insert the AF control board ² into the BX-UCB by aligning the orientation
of the connector inside the BX-UCB and that of the AF control board.
Push in the board along the board rails all the way so that the connector
is plugged in securely.
}A space of two board rails is required for installation. Use two adjacent
option slots on either the central or right.
3. Clamp the AF control board ² using the clamping knobs removed above.
Also attach the other removed cover ³ using the other clamping knobs
removed above.
}Store the other, unused option board cover in a safe place.
3
Assembling Other Modules
Mount all of the applicable modules selected by the user.
At the end, confirm that the main switches of the connected modules are
set to "
" (OFF) and connect the cables.
"