4.5. Calibration for sound alarming range
As shown in the illustration to the right, place the unit on a level
platform and observe the position of the alarm rod. For equal
alarming angles in all directions, the rod should be centered between
the contact ring. The ring can be adjusted by reheating the solder to a
molten state and repositioning the ring accordingly.

5.0. Troubleshooting Guide

No. Symptom
1
Unable to power on (no laser is
emitting, extinguished power
2
indicator light, no sound alarm if
3
titled beyond tolerance)
4
5
After power on, the indicator is not
lighted
6
7
No laser is emitting
8
9
10
No sound if tilted beyond tolerance
11
12
No alarm if tilted beyond tolerance
Prepared By: Tim Wojo
Cause
Dead batteries
Disconnected power line
Switch is broken.
Main circuit board is broken.
The indicator is broken
Main circuit board is broken
Hairspring is cut off
Laser source is broken
Main circuit board is broken
Beeper is broken
Main circuit board is broken
Main circuit board is broken
Revision - Date: 0 – 20080430
40-6680 Service Manual
Repair method
Replace with new battery (3*AA)
Weld the leading wire or replace the
battery case
Replace the switch
Replace the main circuit board
Replace the indicator
Replace the main circuit board
Replace the hairspring
Replace the laser source
Replace the main circuit board
Replace the beeper
Replace the main circuit board
Replace the main circuit board
Page 16 of 16