- ページ 11
温度計 AccuMed AT2103のPDF マニュアルをオンラインで閲覧またはダウンロードできます。AccuMed AT2103 14 ページ。 Non-contact infrared thermometer
-Instructions for recalibration:
While the device is On, Press the "MODE" and "°C/ °F" buttons simultaneously for
3 seconds.
The screen will then display : "F-4"
Press "MODE" to increase by 0.1°F,
Press "°C/ °F" to reduce 0.1°F.
Press "MEM" button to save the setting.
*In the cases of seasonal or environmental changes, a verification and adjustment should be
carried out.
7. CHANGING THE BATTERIES
Display: when the LCD screen displays the flashing symbol "
needs to be replaced soon
Operation: Open the lid and change the batteries, taking great care with the correct
positioning. A mistake with this could cause damage to the apparatus and compromise the
guarantee of your AT2103. Never use rechargeable batteries. Use only batteries for single usage.
IX. TECHNICAL SPECIFICATIONS
1.Normal using condition
Ambient temperature: 50°F ~ 104°F (10°C ~ 40°C)
Relative humidity: ≤85%
2.Batteries: 2 x AA batteries (Included)
3.Unit size: 6.9 x 1.9 x 2 in (L x W x H)
4.Unit weight (including batteries): 0.32lbs (148g)
5.Temperature display resolution: 0.1°F (0.1°C)
6.Measuring range:
In "Body" mode: 89.6°F ~ 109.2°F (32°C ~ 42.9°C)
In "Body" mode, there are three color-identifiers for temperatures:
Green color backlit: ≤99.1°F (37.3°C), means normal temperature.
", the battery is low and