- ページ 6

医療機器 AccuMed CMS-50DのPDF マニュアルをオンラインで閲覧またはダウンロードできます。AccuMed CMS-50D 14 ページ。 Fingertip pulse oximeter
AccuMed CMS-50D にも: ユーザーマニュアル (2 ページ)

b)Humidity: 35%~75%
c)Pressure: 700hPa~1060hPa

3. Principle and Caution

3.1. Principle of measurement

The measurement of pulse oximeter is that it uses a multi-functional oxyhemoglobinometer to
transmit some narrow spectrum light bands through blood samples, and to measure attenuation of
spectrum with different wavelengths according to the characteristic that RHb, O
COHb absorb the light of different wavelength, thereby determining O
Hb saturation is called "fractional" O
fractions. O
2
Fractional O
Hb saturation = [O
2
Oppositely, pulse oxygen oximeter measures functional O
Functional O
Hb saturation = [O
2
Present SpO
oximeter transmits light of two wavelengths only, red light (wavelength 660 nm) and
2
infrared (wavelength 940nm), to differentiate HbO
LEDs, and the other side contains a photoelectric detector. SpO
in the blood by the light plethysmograph when the pulse beats. The result is quite precise when
HbO
saturation is over 70 % ~ 100%.
2

3.2. The attention of Operation

A. The equipment should be fully tested to see if it can be used normally before using.
B.
The finger should be placed properly (see the attached illustration of this manual), or else it
may cause inaccurate measurement.
C.
The SpO
sensor and photoelectric receiving tube should be arranged in a way with the testee' s
2
arteriole in a position in between.
Hb saturation.
2
Hb / (RHb + O
Hb + Met Hb + COHb) x 100
2
2
Hb / (RHb + O
Hb)] x 100
2
2
from HbR. One side of the sensor contains two
2
Figure1. Operating Principle
3
Hb saturation of different
2
Hb saturation:
2
oximeter measures HbO
2
Hb, Met Hb and
2
saturation
2