- ページ 15

血圧モニター AccuMed ABP802のPDF 取扱説明書をオンラインで閲覧またはダウンロードできます。AccuMed ABP802 16 ページ。 Upper arm blood pressure monitor

(>95 % dip in U ) T
for 5s
Power
3 A/m
frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
NOTE UT is the a.c. main voltage prior to application of the test level
Guidance and manufacturer's declaration – electromagnetic immunity
The MODEL ABP802 is intended for use in the electromagnetic environment specified below. The
customer or the user of the MODEL ABP802 should assure that it is used in such an environment.
Immunity test
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by
(>95 % dip in U ) T
for 5s
3 A/m
IEC 60601 test
Compliance
level
level
3 Vrms
3 Vrms
150 kHz to
80 MHz
3 V/m
3 V/m
80 MHz to
2.5 GHz
Power frequency magnetic fields should
be at levels characteristic of a typical
location in a typical commercial or
hospital environment.
Electromagnetic
environment--guidance
Portable and mobile RF communications
equipment should be used no closer to
any
part of the MODEL ABP802, including
cables, than the recommended separation
distance calculated from the equation
applicable to the frequency of the
transmitter.
Recommended separation distance
where P is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter manufacturer
and d is the recommended separation
Distance in meters (m).
Field strengths from fixed RF transmitters,
as determined by an electromagnetic site
survey, a should be less than the
compliance level in each frequency range
b.
Interference may occur in the vicinity of
equipment marked with the following
symbol:
13
80 MHz to 800 MHz
80 MHz to 2,5 MHz