Operation Manual
Revision History
May 2008 – First Release
January 2010–Rev 01
Notices
Shipping Address: Clinical Dynamics of CT, L.L.C.
10 Capital Drive
Wallingford, CT 06492 U.S.A
Sales & Service:
800.247.6427 (toll free in the U.S.)
Sales E-Mail:
Service E-Mail:
Tech Support:
203.269.0090
Support Email:
Facsimile:
203.269.3402
Web site:
www.clinicaldynamics.com
Trademarks
AccuPulse™, CalTables Technology™ and CalTables™ are trademarks of Clinical
Dynamics of CT, L.L.C.
Patents
The AccuPulse is protected by U.S. Patent 5,027,641.
Restrictions and Liabilities
The information contained in this manual is subject to change. Any changes made to the
information in this manual will be included in future manuals. No responsibility is
assumed by Clinical Dynamics for the use or reliability of calibrations, software upgrades
or equipment not supplied by Clinical Dynamics or its authorized dealers and authorized
service centers.
The contents of this document, including all figures and drawings, are proprietary
information of Clinical Dynamics of CT, L.L.C., provided solely for purposes of operation,
maintenance or repair, and dissemination for other purposes or publication, or copying
thereof is prohibited without prior written consent by Clinical Dynamics of CT, L.L.C.,
Wallingford, Connecticut.
In the interest of continued product improvement, all specifications are subject to
change without notice.
All Rights Reserved
Copyright © 2009, Clinical Dynamics of CT, L.L.C.. All rights reserved.
No part of this manual may be transmitted, reproduced, translated,
transcribed copied or electronically transmitted without prior written
permission of Clinical Dynamics of CT, L.L.C.. Printed in U.S.A.
Warranty Service & Shipping Instructions
Warranty Service
All repairs on products under warranty must be performed or approved in writing by
Clinical Dynamics Service personnel. Unauthorized repairs will void the warranty.
Assistance
If the product fails to function properly, or if assistance, service or spare parts are
required, contact Clinical Dynamics Customer Service at 800-247-6427 or visit our web
site, www.clinicaldynamics.com and select the SERVICE & SUPPORT MENU. Select
TECHNICAL SUPPORT REQUEST and complete the form. After this form has been
submitted, a Clinical Dynamics' Service Representative will contact you to help solve
your problem. Before contacting Clinical Dynamics, please attempt to duplicate the
problem and to check all accessories to ensure that they are not the cause of the
problem. Prior to calling, please be prepared to provide:
User Manual
AccuPulse Plus – Hand Held NIBP Simulator
(v1.3.x and v2.x.x software)
Operation Manual
AccuPulse™ is a trademark of Clinical Dynamics of CT, L.L.C.
Clinical Dynamics of CT, L.L.C. 10 Capital Drive • Wallingford, CT 06492 U.S.A.
www.clinicaldynamics.com Phone: 203.269.0090 • Facsimile: 203.269.3402
1
• 1. Product name, model number and serial number.
• 2. A complete description of the problem including the conditions under which the
problem occurred. Ideally, a written problem description would be provided, allowing
for more efficient handling of your initial service request and the subsequent
diagnosis and remedy of the problem.
• 3. Your institution's complete name and address with a contact name and phone
number.
• 4. A purchase order number if the product needs non-warranty service or if you are
ordering spare parts.
Returning a Product for Service
Contact Clinical Dynamics Customer Service at 800 247-6427 and provide the
information
listed
above
under
www.clinicaldynamics.com and select the SERVICE & SUPPORT MENU. Select
CALIBRATION / SERVICE REQUEST to fill out a Request for Calibration/Service Form. It
is highly recommended that you pack the product in its original shipping carton and
packing materials, provided that they are still usable. If the original packaging is not
available, select a sturdy corrugated carton large enough to hold whatever items you are
returning, and also to allow 4 to 6 inches of packing material on all sides of the items.
Whether you use the original packaging or an appropriate substitute, please follow these
packing instructions:
1. Remove all hoses, cables, power cords and any other accessories from the instrument
and, if possible, place the instrument is a clean plastic bag. Note:if you are using
substitute packaging, it is essential that you seal the instrument in a clean, static free
plastic bag or in clean bubble wrap in order to prevent packing material from entering
the product.
a)2. Pack only the accessories you are requested to return; place them in a separate bag.
b)3. If you are using substitute packaging, create a foundation of 4 to 6inches of packing
material (either bubble wrap or packing "peanuts") at the bottom of the carton.
c)4. Insert the instrument and the accessory bag into the shipping carton.
d)5. If you are using substitute packaging, fill the 4 sides and the top of the carton with
4 to 6 inches of packing material (either bubble wrap or packing "peanuts"). Ensure
that the instrument and accessory bag are held firmly in place by the packing
material.
e)6. Please place paperwork such as the purchase order, contact info and reasons for
return in the top of the carton.
f) 7. Close the carton and securely seal it with tape; since in most cases the carton will
have been previously used, it may be necessary to reinforce the original tape on the
bottom of the carton.
g) 8. Ship the product via whatever carrier (UPS, FedEx, etc.) is most convenient.
However, please be aware that, depending on where you are shipping from, standard
UPS shipping could take as long as 7 business days. Unless other arrangements are
made, Clinical Dynamics will return the repaired product to you via UPS.
h) 9. Shipping insurance is optional. Claims for damage to the product during shipping
must be initiated by the shipper.
Additional Instructions for International Shipments
Customers outside the United States must include a "pro-forma invoice" for customs
purposes as part of their shipping documents. It is imperative that the name of the
product appears exactly as follows:
AccuPulse Test Equipment
The use of any other product name could add unnecessary delays when shipping
internationally.
Assistance,
or
visit
our
web
site,