- ページ 9
アクセサリー AccuTherm ACCSL2021のPDF マニュアルをオンラインで閲覧またはダウンロードできます。AccuTherm ACCSL2021 12 ページ。
AccuTherm ACCSL2021 にも: マニュアル (15 ページ)
SPECIFICATIONS:
SmartLOG® DDL Monitor Gateway
Wi-Fi:
802.11 b/g/n (2.4 GHz only)
Security:
WPA/WPA2 Enterprise
Display:
240x320 pixel TFT Color Display w/ Touch Screen
Screen Update:
60 seconds
Memory:
16MB (2 years of records @ 5 minute intervals)
Ambient Sensor:
Temperature range -40 to +125°C (-40°F to 257°F)
Humidity range 0-100%RH
Alarms:
Audible Piezo Buzzer / Visual LED / SMS / Email
Date/Time:
NTP clock synchronization during boot
Power:
5v wall adapter w/ 5' micro USB cable
Batteries:
Alkaline 1.5v 3 x AA (backup backup only for power loss)
~ 3 days with no alarms
Battery Life:
(NOTE: replace all batteries after any power loss of more than 1 day)
Working Temp:
-20 °C ~ 70 °C
Dimensions:
83 (W) X 120 (H) X 26 (D) mm
Wireless Buffered Temperature Sensor
Temperature range:
-40C to 50°C (-40°F to 122°F)
Accuracy:
±0.5°C (±1.0°F)
Distance:
50m/164ft (open field), 10m/32ft (inside fridge/freezer)
Buffer medium:
Glycol
Batteries:
Alkaline 1.5v 2 x AAA
Battery Life:
~1yr
Certification:
2yr NIST ISO17025 Certificate Included
SETTINGS PAGE Access Pop-up
Pin: 5 digit number
If created during initial setup enter Pin number,
otherwize, leave blank and click Open.
SETTINGS PAGE
--Wi-Fi Setting--
These fields are populated from the initial
SmartLOG SETUP.
To change network...
Network Name: SSID (Service Set Identifier)
is the name of your wireless network
Password: network password.
Username: is for Enterprise Networks.
--Logger Configuration--
Access Pin: 5 digit number (required to save)
Logger Name: The name is used to access the
Dashboard and also for Email & SMS messages
sent. Do not use the same name if using
additional data loggers on the same network. 12
alphanumeric characters, hyphen "-", no spaces.
Facility Name:, Location:, Pin or ID#:
are optional fields that show on reports.
P1 & P2 Name: A drop down list to select Fridge
or Freeze that will auto populate Alarm Temp:
min and max values. Custom name, min/max
values can be entered. For single sensor set up,
leave P2 fields empty.
Sensor SN#: Enter serial number of orginal
wireless sensor or new replacement sensor.
Cert Expires: Add 2 years from in service date.
Scale: Display & Report in °C or °F.
Log Interval: Select from 1 to 60 minutes.
Time Zone: Set your local Time Zone.
Time is automatically synchronized from internert.
Note: All times are in Standard Time, no DST.
--Email From Acct Info--
see note 'Office 365 accounts'
Email: Sign in to Email for sending alerts
Password: Email account password
Outgoing SMTP: Dropdown list of popular email
provider servers. Selecting a server name will
auto populate the Port #:
Custom server address & port # can be entered.
If Alerts and Status and message are not needed,
leave Email From Acct fields empty.
--Email / SMS Contacts--
Email: Address to send alert email to.
SMS: 10 digit phone number to send text msg to.
Carrier: Select carrier of cell phone #.
Select from dropdown list of popular cell phone
carriers or enter custom value.
--Daily Status Messages--
Time: can be set to OFF for both am and pm to
disable status messages.
Select one or both am or pm for desired
notification times.