IX. SYSTEM TEST (CONDUCTED BY INSTALLER PRIOR TO USE)
1. Turn system on by depressing the on/off button on the wall or desk alarm.
2. Press the test button - all four lights should blink and the audible alarm should sound. To turn off
audible alarm press cancel button.
X. TROUBLESHOOTING
Problem
High/low pressure alarm doesn't work
Possible solutions
• Pressure switches not plugged in securely at manifold
• Pressure switch (O 2 and/or N 2 O) requires adjustment
• Gray cable run from the wall or desk alarm was damaged during construction
Problem
No lights on wall or desk alarm
Possible solutions
• Power supply not plugged into 120-volt electrical outlet
• Power supply not plugged into manifold
• 120-volt outlet is not receiving power
• Gray cable is not plugged into the manifold or the wall or desk alarm securely