Siemens SIREMOBIL Compact L 取扱説明書
Replacement of the 15" LCD Color TFT Display

Safety Information

WARNING
WARNING
WARNING
CAUTION
CAUTION
Siemens AG
Medical Solutions
Risk of physical injury or death and property damage!
Observe
- the product-specific safety information in these instructions
and the service instructions of the product,
- the general safety information in the instructions
TD00-000.860.01... and
- the general safety information according to ARTD Part 2.
Noncompliance can lead to death, physical injury or property
damage.
Electrical voltage
Certain components inside the TFT Display are under
high voltage!
Do not open the TFT Display housing! This is not necessary in a
service situation.
Do not insert objects into the TFT Display housing!
Electrical voltage!
Noncompliance can lead to death, physical injury or property
damage.
After completing all service work and attaching all system covers,
you must perform a protective ground wire test according to
ARTD-002.731.17... .
The protective ground wire resistance must not exceed 0.2 ohms.
Risk of being burned by hot parts or components!
Noncompliance can lead to light or medium burns, especially of
the hands.
Once the covers have been removed, parts or components (e.g.
power components, heat sinks, solenoid brakes) become acces-
sible which may exhibit temperatures of up to 50°C in operation.
To prevent burns, switch off the system power supply before con-
tacting any parts or components and let the system cool off for at
least 5 minutes before resuming work.
Electrical voltage!
Noncompliance can lead to property damage.
ESD protection guidelines must be observed when servicing.