General Information
1.6.3
Radiation safety information
WARNING
1.6.4
Mechanical safety information
CAUTION
Siemens AG
Medical Solutions
X-ray radiation!
Non-compliance can lead to illness, irreversible damage to body
cells and the genotype, severe injury and even death.
When performing work on the system during which radiation must
be released, the radiation protection directives and the rules for ra-
diation protection according to ARTD 02.731.02 must be complied
with.
Please note:
Use available radiation protection devices.
Wear radiation protection clothing (lead apron).
Stay as far away as possible from the radiation source.
Release radiation only if necessary.
Set the radiation activity as low as possible. (low kV and
mA values, short radiation time)
Release radiation for as short a time as possible.
Checks in which radiation must be released are identified
by the radiation warning symbol.
Risk of burns from hot parts or components! If not observed, mi-
nor to more severe burns, especially on the hands, can occur.
Parts and components (e.g., power components, cooling fins,
electromagnetic brakes) that can exceed 50 degrees Celsius dur-
ing operation are accessible after the covers are opened.
To avoid burns, switch the system off prior to touching
parts or components and allow at least 5 minutes of cool-
ing.
SPR2-230.101.02.06.02
07.05
Page 13 of 46
CS PS 24
13
SIREMOBIL Iso-C