2.7
Operating values - inspection
2.7.1
Lifting base
PMF
Lifting drive chain
•
Check the tension on the lifting chain.
- Raise the table all the way up.
- Pull the lifting drive chain with the spring scale approx. 5mm toward the cover (Fig. 5).
When the lifting drive chain defelects 5mm, you must read 60
2.7.2
Spotfilm device
QSQ
Radiation field
•
Check the coincidence of the light and radiation fields with the large focus.
- Put a 24 x 18 cassette loaded with film on the tabletop and mark the edge of the light
field with washers.
- Release an exposure using values 45 kV, 3.2 mA.
- Develop the film.
- Measure the deviations between the light and radiation fields; they must be smaller than 0.5 cm.
UROSKOP D
Inspection and maintenance
cover
RLL5-310.101.02
Page 12 of 16
Rev. 02
01.97
TDSP 1
±