reference electrode against different acidity of solution to be measured. The D.C. potential is fed to an
analog-to digital converter via a preamplifier for digital display of pH value. Similarly, in an application
that it is equipped with an appropriate ion selective electrode for a potential titration, it is need for
display of terminus potentials.
1. Measuring theory
Ⅳ. Structure
The meter make up of electro-unit, compound electrode and temperature sensor as Fig 2.
Ⅴ. Operation and Maintenance
1. Setting the meter
The meter must be set on no vibration table and have good earthing.
2. Preparation before using
2.1 Warm up the meter
a) Take out the compound electrode and temperature sensor. Preparing buffer solution, distilled water,
beaker and filter paper etc. The Q9 short circuit plug must be inserted into the input terminal of the
instrument before inserting the electrode into the instrument, to make the input terminal in the short
circuit and to protect the electric element.
b) Put the power key on "OFF". Insert the three-pin power plug into the AC power outlet.
c) Put on the power supply. Warm up the meter for 10 minutes before using it.
2.2 How to use the ATC and MTC function.
To link the temperature sensor, put the selection knob on auto position and the meter can work in ATC
mode. Now the MTC function will be invalidation.
Fig. 2 The structure of pHS-3B
1.
Temperature adjusting knob
2.
Slope adjusting knob
3.
Setting knob
4.
Selection knob
5.
Power outlet
6.
Power switch
7.
Fuse seat
8.
Connection
terminal
electrode
9.
Measuring electrode outlet
10. Hole for electrode rod
11. Indicator light
12. Selection switch
13. Temperature sensor
for
reference