IR Link Software User Guide:
8.3. Software Installation
Up to date software is available from your local service agent,
Crowcon office. This software will work with Windows versions XP, Vista, and Windows 7.
8.4. Getting started
Below are instructions on how to use the IR Link software. Once
you have successfully installed the software and click on the IR
Link icon, you will see the screen to the right:
In order to interact with your Clip & Clip+ make sure to have the IR
Link plugged into your computer. Place the monitor face down at
least 2 inches in front of the IR Reader, with the top side facing the
IR Reader. Then click "Read Device" if the read was successful
you will see a green bar at the bottom of the screen.
8.5. Changing Settings
Both the Clip & Clip+ monitors allow you to change the following
fields, below is a brief description of each field and changes
possible:
a. High Alarm/Low Alarm Set Points: Adjust the high/low alarm
set point. Note: When doing so make sure to set the alarm
limit not to exceed the local or site guidelines. Some sites
may have different set points than others.
b. Bump Interval: In order to help with bump test accountability users can adjust the Bump Interval.
If a unit has not been bump tested within the time frame the unit will emit a flash until it has been
tested either manually or via the C-Test Station.
c. User ID: Each individual unit can now have a unique electronic user ID. You are able to program
in up to 6 characters. Note: Some alphabetic characters do not display well.
d. Self Test Interval: By default all Clip & Clip+ monitors have a default 20 hour self test. Frequency
can be set anywhere from 8 to 20 hours.
e. Show Sensor Readings: Once the unit starts to see gas the unit will display the real time reading
with a 0.1 ppm resolution. If this is unchecked the unit will always display the life remaining or
flash low/high alarm in an alarm condition
f. Hide bump due LED Flashing: If you prefer to not have the units flash when due for a bump test,
please make sure to check this box.
g. Hide Clock: This will turn the monitor into a real time ppm display monitor. By default always
showing a real time ppm reading.