- ページ 2
計測機器 AEMC instruments SR704のPDF ユーザーマニュアルをオンラインで閲覧またはダウンロードできます。AEMC instruments SR704 4 ページ。 Ac current probe

CURRENT PROBE - SR701/SR704 DRAWING
ELECTRICAL SPECIFICATIONS
Current Range: 1 mA to 1200 Aac
Transformation Ratio: 1000:1
Output Signal: 1 mA AC/Aac
(1 Aac at 1000 A)
Accuracy and Phase Shift*:
Accuracy:
(1 to 100) mA:
±3 % Reading ±5 mA
(0.1 to 1) A:
±2 % Reading ±3 mA
(1 to 10) A:
±1 % Reading
(10 to 100) A:
±0.5 % Reading
(100 to 1200) A: ±0.3 % Reading
Phase Shift:
(1 to 100) mA: Not Specified
(0.1 to 1) A:
Not Specified
(1 to 10) A:
≤ 2 °
(10 to 100) A: ≤ 1 °
(100 to 1200): ≤ 0.7 °
*Referenced conditions: 23 °C ± 3° K, (20 to
75) % RH, (48 to 65) Hz, external magnetic
field < 40 A/m, no DC component, no external
current carrying conductor, test sample cen-
tered. Load Impedence: ≤ 1 Ω
Overload: 1200 A for 40 min on, 20 min off
Accuracy: Per IEC 185
Frequency Range: 30 Hz to 5 kHz; current
derating above 1 kHz using the formula:
1000 A ×
1
F (kHz)
Load Impedence: 5 Ω max
Working Voltage: 600 V CAT III
Common Mode Voltage: 600 V CAT III
Open Secondary Voltage:
< 25 V by limiting circuit
Influence of Adjacent Conductor:
< 1 mA/A
AC