12) Press
F1
to enter the Print rate
number entry screen. The current
print rate appears on the LCD.
In this example, we'll set the Print Rate
to an output rate of once every 2
seconds.
13)
PressF2
to change the Print rate.
Use F2 to scroll the number,
enter each number with the F1
key.
14) When the entire number is
displayed, press
finalize the seconds entry. The
next Print Menu Item appears,
"String".
15) Exit the Print Setup Menu by
pressing
ZERO
message "Store" appears briefly
then normal system operation
starts.

5.9 COMM Port Hardware

The MSI-8000 RS-232 Comm Port is used for software updates, connecting to a remote
display and for connecting to any RS-232 device.
Connector: M12 industrial IP67 rated. An adapter cable (MSI P/N 503489) is required to
connect the MSI-8000 to a computer. This adapter cable converts the 8000 connector to a
standard D9 serial connector. The 503489 cable can be converted to DTE by using a Null
Modem adaptor.
Data Configuration: The data output is fixed at 8-1-N.
Baud Rate: Programmable for 300 to 230.4<baud in 8 steps. The bootloader for updating
software is always 38.4k baud
Handshaking: No hardware handshaking is supported. Xon/Xoff software handshaking is
always on.
It may be necessary to disconnect the shield drain wire at the D-9 connector frame to
prevent ground loops. Ground loops can cause unstable readings. In extreme cases it may
be necessary to use an isolated RS-232 interface.
Serial Cable Schematic, DCE Configuration for connecting to a computer (503489-0001
schematic).
28
MSI-8000
Operator's Manual
ENTER/SELECT
ENTER/SELECT
ENTER/SELECT
F1
again to
EXIT/SAVE
twice. The
Figure 5.6 Printer/Output Set Up (continued)
blinking
Example Value
set once per hour, set the rate to 3600, etc. Largest possible
entry is 65535 which is >18 hours.
SCROLL
blinking
fixed
0
ZERO
The number represents the print
interval rate in seconds. Use 0 for the
fastest the MSI-8000 can go (~5 times
per second). To set a once a minute
transmission, set the rate to 60. To
SCROLL
blinking
EXIT/SAVE
0
ZERO