REF: Druglizer LE5 Drug Tester User Manual V1
Technology
• Approximately 500uL (half a millilitre) of
saliva is required per drug test.
• If insufficient saliva is taken, an 'Invalid'
test result may occur.
• A Collector Indicator is located opposite
the Collector pad and will change to a blue
colour when sufficient saliva is taken.
Firstly, instruct the subject to insert the
collector pad into their mouth then:
• rub on inside of both cheeks;
• collect saliva around mouth and
under the tongue.
Ensure the sponge is not chewed.
• Instruct the subject to insert the Collector
into their mouth and rub the pad onto the
inside of both cheeks as well as under the
tongue and around the mouth.
• IMPORTANT: Druglizer LE5 training MUST
be undertaken in order to ascertain correct
saliva collection. Collect saliva until the
Collector indicator changes to a BLUE that
is consistent from edge to edge.
• When enough saliva is taken, the Collector
Indicator will change to a BLUE that is
consistent from edge to edge
Page 10
©