- ページ 3

コントロールユニット Alcolizer Druglizer LE5のPDF ユーザーマニュアルをオンラインで閲覧またはダウンロードできます。Alcolizer Druglizer LE5 16 ページ。 Charging cradle
Alcolizer Druglizer LE5 にも: ユーザーマニュアル (20 ページ), ユーザーマニュアル (19 ページ), 組み立て説明書 (3 ページ)

Alcolizer Druglizer LE5 ユーザーマニュアル

1 INTRODUCTION

The Druglizer LE5 Drug Testing System is for the rapid oral fluid screening of drugs of abuse.
The system incorporates a Druglizer LE5 Drug Tester Cartridge (Druglizer LE5 Cartridge), a
Druglizer Oral Fluid Collector and a Druglizer LE5 Drug Tester instrument (Druglizer LE5,
purchased separately). The system is intended to be used by a trained professional to provide
a first line screen. A positive result displaying 'Unconfirmed' on-screen will require
confirmation by a laboratory-based reference assay like Gas chromatography – Mass
Spectrometry (GCMS) or Liquid Chromatography Mass Spectrometry (LCMS).
This document must be read carefully before using the Druglizer LE5 Drug Testing System and
Certified Operator Training provided by Alcolizer Technology must be completed before using
the Druglizer LE5 Drug Testing System. Keep the user manual available for the life of the
instrument and ensure operators have access to the user manual at all times. You must always
hand the user manual over to the subsequent instrument owner or user.
WARNING
• Only human oral fluid should be used.
• Results on the Druglizer LE5 are preliminary qualitative analytical test results. A
result displaying 'Unconfirmed' on-screen will require confirmation by a laboratory-
based reference assay.
• The presumptive presence of drugs is not a measure of impairment.
• A negative result may not necessary indicate a drug free sample. Drugs may be
present below the stated cut off.
• Foods/medicines or preclinical errors may result in false results.
The Druglizer LE5 must be stored at a temperature between –10°C and +60°C. Conduct drug
screening tests at a temperature between +5°C and +35°C.
This User Manual is applicable to the Druglizer LE5 with main module firmware version RN-1.5.0

2 TEST PRINCIPLE

The Druglizer LE5 Drug Testing System is based on a competitive lateral flow immunoassay
technique where drug molecules that may be present in the oral fluid sample compete with
the same drug attached to the nitrocellulose wicking membrane, for limited binding sites on
tagged drug specific antibodies.
The antibodies are tagged with gold colloids to produce a conjugate with a red colour. During
testing, an oral fluid sample is absorbed onto the oral fluid collector pad. Chemicals react with
the saliva to remove sample interferences and a portion of this sample wicks onto the
REF: Druglizer LE5 Drug Tester User Manual V1
NOTE
and drug test module or later
Page 3
© Alcolizer Technology