• On the press passive test page, press the Test Key. The instrument will perform a
background self-test while Please wait....is displayed on the screen.
• The instrument will then display the Ready for passive test screen (Refer Figure 14).
• Hold the instrument no further than 5cmin front of the test subject's mouth; so that
the subject is breathing or talking directly into the instrument's sample inlet port
(Refer Figure 15).
• Ask the subject to start counting from 20 onwards in a one second rhythm, until you
tell them to stop,"20...21...22...23...24...25...etc."
• Press the Test button while the subject is counting. The instrument will make a series
of 3 clicking sounds (pump clicks) as the sample is being taken.
• The passive test will have 3 short beeps and clicks followed by a longer beep.
Position the mouth 5cm away from the breath entry port. Blow steadily into the
passive cup, listening until a long beep is heard indicating the breath collection is
complete. Stop blowing. The instrument will take a short moment to analyse the
sample followed by displaying the result.
During these three pump activations, the operator must hold the instrument steady and at
the same distance from the subject. After the three pump clicks, the instrument will sound a
long beep indicating the test is complete. Analysing Sample will then be displayed (Refer
Figure 11).
3.5.7
Analysing a Passive Breath Sample
After a valid breath sample is taken, the Analysing Sample message is displayed until the
instrument determines if alcohol is present. This will normally take a few seconds. In cold
climate conditions it may take longer to display the result.
When the Passive standard reading is calculated, it will be displayed and retained until the
recovering period has ended.
REF: Alcolizer LE5 Alcohol Tester User Manual V15 docx
Figure 14- Ready for Passive Test Message
Figure 15- Providing a Passive Sample
Page 16
© Alcolizer Technology