試験装置 Alcoscan AL3500のPDF 取扱説明書をオンラインで閲覧またはダウンロードできます。Alcoscan AL3500 19 ページ。 Coin-operated breath alcohol tester

Alcohol Level - Low Limit
This setting adjusts the detected alcohol level below
which the AL3500 will light the "Good" Indicator after a
test is performed.
The default setting is 0.02. We do not
recommend changing this setting.
Tap SET1 until you see "Lo" in the LED display. Tap
SET2 to adjust the limit from
Alcohol Level - High Limit
This setting adjusts the detected alcohol level above
which the AL3500 will light the "Danger" Indicator after a
test is performed.
The default setting is 0.05. We do not
recommend changing this setting.
Tap SET1 until you see "Hi" in the LED display. Tap
SET2 to adjust the limit from
0.00
to 0.04.
0.05
to 0.10.
12