- ページ 7
試験装置 Alcovisor JupiterのPDF 取扱説明書をオンラインで閲覧またはダウンロードできます。Alcovisor Jupiter 12 ページ。
Alcovisor Jupiter にも: マニュアル (8 ページ), 取扱説明書 (12 ページ)
OPERATING INSTRUCTIONS
Pre-test Conditions
1. Operating temperature range is -10°C to 50°C. Use within this temperature
range will limit condensation from a subject breath to insure an accurate
reading from a breath sample.
2. It is recommended that the operator to observe a subject for about 20 minutes
before administering the breath test to insure that nothing is consumed or
expelled from the mouth of the subject.
eliminate reading from mouth alcohol.
Standard Test
1. Device takes a blank test before each test to prevent residue. Wait for 30
seconds and start the test again if the blank test result is not zero.
2. Input information by a stylus pen.
NOTE: Maximum 16 characters for each input.
3. When displays "Please Blow", attach a disposable mouthpiece onto the
sample port, take a deep breath and blow until BEEP tone stops.
4. Test result displays on the display.
5. Select
to print the test result.
6. Select
7. If testee did not blow within 30 seconds, tester can select "Refuse" to confirm
the testee refused to do the test; or select "Test Again" to allow testee to do
the test again.
8. If testee failed to give a valid breath sample, tester can select "Discontinue" to
confirm; or select "Test Again" to allow testee to do the test again.
9. A passive test can be performed by select "Manual" during the testee is
blowing.
10. Push the button underneath the mouthpiece to remove the mouthpiece.
11. 65,000 test records can be stored, if memory limit is reached, new test is not
available. Send the device to service centre to upload test results to PC,
and clear out all memories.
to start another test.
Following this procedure will
P. 5