Start-up
1. Switch on the numbered components in order and refer to figures 5-10 for details of the controls.
2. Using the joystick, move the stage (fig. 9) so that the objective is centrally positioned. The speed of the XY-
stage can be varied via the XY button above the joystick. A focus wheel can be found on the side of the joystick
unit and the speed of focussing can be changed by pressing Z button on the microscope or joystick. The speed
of movement is indicated on the joystick display panel - ^ denoting slow movement.
3. Objective Lenses on the live cell system are 4x, 10x, 20x and 40x dry objectives. Check the correction collar is
set to 0.17 on the 20x and 40x objective lenses for imaging through standard #1.5 coverglass; the collar can
be adjusted for thicker samples. 60x and 100x objectives can be added to the system but are not permanently
installed.