table. If the magnification out of the
specified range is used, uneven
illumination will occur or the seeing of
the sample image will be degraded.
Mount the P-ICI2 onto the zooming body
1
aligning the groove on the bottom of the
P-ICI2 main body with the positioning pin on
the zooming body while tilting the P-ICI2
toward the rear. Tighten the eyepiece tube
clamp screw.
Note) When the other intermediate tube such
as a beam splitter is used together,
attach it on the P-ICI2.
Screw in the 1/4 lambda plate to the end of
2
the objective. When the C-FIR ring fiber light
guide is used together, use the other 1/4
lambda plate separately available for ring
fiber illuminator.
Insert the incident section of the light guide
3
into the light guide mount of the fiber
illuminator and tighten the clamp screw.
Insert the luminous section of the light guide
into the light guide mount of the P-ICI2 and
tighten the clamp screw.
30
Groove
Positioning pin
Luminous
section
1/4 lambda plate