TEST PROGRAM
• Adjusting the cylindrical axis
Using the [INCR] and [DCR] keys, adjust the cylindrical axis displayed in the
" AXIS=" field matches the cylindrical axis value displayed on the mounting holder
of J38056 astigmatism model eye tool.
[INCR] key: Increases the cylindrical axis by one. The value displayed in the
"OFFSET^" field is also increased.
[DCR] key : Decreases the cylindrical axis by one. The value displayed in the
"OFFSET=" field is also decreased.
* * * Set a cylindrical axis sign to +C.
• Saving the offset value
When the offset value displayed in the "OFFSET=" field simultaneously during the
cylindrical axis adjustment is established, set DIP switch No. 5 to ON and press the
[SET] key.
The offset value will then be stored in the EEPROM and two short beeps will be
generated. Then reset DIP switch No. 5 to OFF.
© TEST -13 : Selection of the model eye mode and human eye mode and parallel
shifting of the spherical power
Press this
key to quit
TEST-13.
• Selection of the model eye mode and human eye mode
[HUMN EYE]
: Must be selected to measure the human eye and model eye
provided with the instrument (normal measurement mqde).
(MODL EYE]
: Must be selected to measure the model eye tools during
measurement accuracy test.
Press the [End] key to return to the refractometer test program
select screen and press the [EXIT] key to enter the normal
measurement mode screen.
The "R/L" display on the normal measurement mode screen
will turn to "M". Now you can measure the model eye tool
accurately.
* After the model eye tool measurement is complete, turn the Main switch off and
on to return to the normal human eye measurement mode.
• Parallel shifting of the spherical power
Use this test to shift the measured spherical power parallel.
[+000] key:
Used to select the shifting amount. Each press of this key changes
the power displayed on the key top.
[6-141
Repair Manual