6 PREVENTIVE MAINTENANCE

6.1 GENERAL

The monitor should be wiped clean with a damp cloth following a regular maintenance program.
Avoid spraying, submersion and other conditions that could cause a liquid to enter the monitor
and cause possible intrinsic damage to internal components.

6.2 SENSOR REPLACEMENT

Caution:
Turn off the power supply before replacing sensor.
The signal from the sensor will be greatly reduced when its replacement is required. The sensor
should be replaced when it no longer responds to the presence of gas or has an unstable zero
signal.
Sensor life as well as the "Sensor Part Number" that is needed to reorder the replacement
sensor are found in the Product Information on page IV.
To replace the sensor:
-
disconnect power
-
remove the four cover screws and cover from the unit
-
unplug the used sensor element from the board and dispose of it properly
-
plug in the replacement element onto the board, making sure sensor pins are
aligned with the pin sockets
-
connect the power perform the calibration procedure
-
following calibration, reinstall the cover plate
refer to Figure 8 for sensor replacement.
-
Note:
Allow 10 minutes for the new sensor to stabilize before performing section 5.2 calibration
procedure.
AMC-1040 Monitor