Model 2500 Quick Start Manual
5
Loosely install the locking collar on each shaft.
6
Install and secure (finger tight) the paddle blades.
7
With vessels in place, gently push down each shaft until the paddle blade touches the
bottom of each vessel.
8
Lock each shaft collar using 5/64 inch hex wrench. Do not over-tighten the set screws on
the shaft collars and avoid putting any weight on the drive head when performing this
step.
For Paddles
1
With the drive head fully raised, position the Quick Height Adjustment block to USP 2
Paddles.
2
Install and secure (finger tight) the paddle blades.
3
Lower the drive head and turn the knob to lock position. The distance of 25±2 mm
between the blade and the inside bottom of the vessel is set and maintained during the
test.
Height adjustment block position for paddles
Height Adjustment
© 2017 Distek, Inc. All Rights Reserved